Measuring Technology for Hard Magnetic Material
——MATS-2010H Hysteresisgraph
◆ Introduction
MATS-2010H Hysteresisgraph adopts ATS structure, users can customize different configurations according to different requirements. Select suitable electromagnet and corresponding test power supply according to the intrinsic coercivity and size of sample; select measuring coil and probe according to different test methods; decide whether select fixture or not according to sample shape; decide whether select temperature controller or not according to test requirements.
◆ Test Items
➲Automatically measure the saturation hysteresis loop and magnetic parameters Js, Br, HcB, HcJ, (BH)max, μrec of general permanent magnets such as Ferrite, magnetic rubber, AlNiCo and so on.
➲Automatically measure the demagnetization curve in the second quadrant and magnetic parameters Br, HcB, HcJ, Hk(H90), (BH)max of rear earth permanent magnets such as bond NdFe, sintered NdFeB, SmCo and so on.
➲Test sample shapes are toroid, round cake, square, tile and other irregular shapes.
➲Select optional tile fixture to directly measure motor tile.
➲Select optional temperature controller to measure the magnetic parameters under different temperatures (50℃ to 250℃).
◆Technical Data
1. Hardware Parameters
1.1 TPS-2X00 power supply
Model |
TPS-2000H |
TPS-2100H |
TPS-2300H |
Input power |
Single-phase 220V 50Hz |
Single-phase 220V 50Hz |
Single-phase 380V 50Hz |
Output voltage |
0~±100V |
0~±180V |
0~±300V |
Output current |
± 20A (Max. Value) |
± 30A (Max. Value) |
± 60A (Max. Value) |
Stability |
Superior to 0.05% |
Superior to 0.05% |
Superior to 0.05% |
Output protection |
Over voltage, over current, short-circuit protection. |
1.2 LZ-210B/H Fluxmeter/Gaussmeter
Model |
LZ-210B/H | LZ-210B2/H |
|
Fluxmeter |
Channel |
Single channel |
Double channels |
Range | 2.5, 5, 10, 20 mWb |
||
Accuracy(full range) | 0.5% |
||
Zero drift | 2μWb/30 sec(2.5mWb) |
||
Gaussmeter |
Range |
0.5, 1, 1.5, 3 T |
|
Accuracy(full range) | 0.5% |
||
Non-linearity | ±0.2%(software correction) |
1.3 DCT Electromagnet
1.3.1 Model and parameters
Model |
DCT130 |
DCT150 |
DCT250 |
Head space |
0~50mm (manual adjustment) |
0~50mm (manual adjustment) |
0~80mm (electric adjustment) |
Pole diameter |
130mm |
150mm |
250mm |
Rated current |
20A |
30A |
60A |
1.3.2 Pole area diameter, air gap and magnetic field
Model |
Pole head |
Pole area diameter(mm) |
Max. Magnetic field (10mm air gap) |
DCT-130 |
Flat pole |
130 |
≥1.5T |
Iron cobalt shrink pole | 60 | ≥2.1T |
|
DCT-150 |
Flat pole |
150 |
≥1.5T |
Iron cobalt shrink pole | 80 | ≥2.5T |
|
DCT-250 |
Iron cobalt shrink pole |
80 |
≥3.0T |
Iron cobalt shrink pole | 120 | ≥2.5T |
1.4 PC-6616 Data Acquisition Card
Parameters |
Specifications |
A/D convert |
16Bit, 8 channels differential input |
Convert time |
< 10μs |
Convert accuracy |
0.01% (full range) |
D/A convert |
16Bit, 2 channels independent output |
Convert accuracy |
0.01% (full range) |
2. Accessories (Optional)
2.1 TCS Temperature controller
Model |
TCS-200 |
TCS-250 |
Max. Output power |
800w |
1200w |
Temperature range |
50~200℃ |
50~250℃ |
Accuracy |
±1°C |
|
Remarks |
Used together with heating pole. |
2.2 Test Poles
Flat pole Iron cobalt shrink pole Tile fixture and test coil
◆System Parameters
1. Working Environment
Parameters |
Specifications |
Input power |
Sigle-phase 220V, 50Hz or three-phase 380V, 50Hz |
Environment |
Temperature: 23±5℃; Humidity: 30~75%RH |
External magnetic field interference |
Absolutely avoid |
Thermal balance time |
10 minutes |
2. System Technical Data
Test AlNiCo reference sample under constant temperature, technical data are as follows:
Measured parameters |
Br(%) |
HcB(%) |
HcJ(%) |
(BH)max(%) |
Uncertainty (k=2) |
1 |
1 |
1 |
2 |
Repeatability (constant temperature) |
± 0.3 |
± 0.3 |
± 0.3 |
± 1 |
◆ HMTest Test Software
1. System and Language
➲HMTest runs in Windows 7 operating system, confirms to Windows software standards, easy to operate.
➲Chinese or English operating interfaces can be selected.
2. Test Functions
➲Automatically set optimal range of excitation voltage, range of fluxmeter and gaussmeter.
➲Magnetization, test and demagnetization of general permanent magnets can be finished at one time, set test time as 20sec~60sec.
➲Rear earth magnets need saturation magnetization before testing, set test time as 60sec~120sec.
➲The sample after testing can be in a state of demagnetization or magnetization, freely selected by users.
➲Fully automatic control, intelligent identification, minimize manual operation.
➲Sampling waveform and instrument status under real-time monitoring, and can be stopped at any time.
➲Automatically complete data calculation and temperature conversion of test results.
3. File and Data Management
➲File system adopts data base, it can directly print or output test results into Excel.
➲File management with powerful functions, it can automatic save data, delete data and delete all data.
➲Data file contains full sampling data, sample parameters and instrument parameters. The file is in text format, and can input to other software.
➲It can display B(H) and J(H) hysteresis loop or magnetization curve, B(BH) magnetic energy-product curve or equal energy-product curve and display coordinate of each point on the curve.
➲Display the curve of many test data at the same time, for example, demagnetization curve or hysteresis loop of one sample under different temperatures; demagnetization curve or hysteresis loop for different samples.
◆ Model Selection
1. Standard Package
Name |
Model and descriptions |
||
MATS-2010H/130 | MATS-2010H/150 | MATS-2010H/250 |
|
Industry computer |
ADVANTECH IPC-510 |
||
Monitor |
LENOVO 18.5' |
||
Software |
HMTest part functions |
HMTest full functions |
HMTest full functions |
Power supply |
TPS-2000H |
TPS-2100H |
TPS-2300H |
Electromagnet |
DCT-130 |
DCT-150 |
DCT-250 |
Flat pole |
130 flat pole |
150 flat pole |
—— |
Iron cobalt shrink pole |
—— |
Φ80mm |
Φ80mm |
Gaussmeter/ Fluxmeter |
LZ-210B/H |
LZ-210B2/H |
LZ-210B2/H |
Data acquisition card |
PC-6616 |
||
Reference sample |
AlNiCo, with the third party test report. |
||
Coil |
B coil or J coil |
B coil, J coil or JH coil |
B coil, J coil or JH coil |
Standard cabinet |
600mm*600mm*900mm |
2. Optional
S/N |
Name |
Model |
Specifications |
Remarks |
1 |
Temperature controller |
TCS-200 |
50~200℃ |
Used together with heating pole. |
2 |
Temperature controller |
TCS-250 |
50~250℃ |
Used together with heating pole. |
3 |
Tile fixture |
Directly test ferrite tile |
Customize according to tile specifications. |
Instillation and training of the instrument
We have three methods for your selection according to your convenient and requirements.
1. We provide installation and operation video along with the goods, customer can refer to the video for installation and operation of the instrument.
2. We provide on-site installation and training service, extra charge is needed.
3. Training on manufacturer site and the training is for free, customer pay all the other travel relative cost.
http://www.hystograph.com/